PicoVNA 108
Oscilloscopes & AnalyzersPicoVNA 108 USB vector network analyzer (VNA), 300 kHz to 8.5 GHz, full 2-port S-parameter measurement with 124 dB dynamic range at 10 Hz and 0.006 dB rms trace noise. This USB-controlled S-parameter analyzer is available in India from GSAS Micro Systems with calibration kits, application engineering and local support.
Frequency
300 kHz to 8.5 GHz
Ports
2-port (S11, S21, S12, S22)
Dynamic Range
124 dB at 10 Hz
Trace Noise
0.006 dB rms (140 kHz BW)
Directivity
47 dB typ / 40 dB min (corrected)
Speed
>5000 dual-port S-parameters/sec
Power
USB-controlled, external 12–15 V DC (25 W)
Software
PicoVNA 5 (Windows 8/10/11 64-bit)
Overview
About PicoVNA 108
The PicoVNA 108 from Pico Technology is a 2-port USB vector network analyzer covering 300 kHz to 8.5 GHz, delivering full S-parameter measurement (S11, S21, S12, S22) with up to 124 dB dynamic range at 10 Hz and 0.006 dB rms trace noise at 140 kHz bandwidth. The instrument connects to and is controlled over USB, and is powered by an included 12 to 15 V DC mains adaptor (25 W), giving benchtop VNA performance in a compact, portable form factor, without the cost and bench space of a traditional rack-mount network analyzer.
As a true S-parameter analyzer, the PicoVNA 108 resolves the insertion loss, return loss, phase, group delay and impedance of RF components, filters, amplifiers, attenuators, cables, connectors and antennas. With over 5000 dual-port S-parameters per second, it is fast enough for tuning on the bench and for production characterization. The 8.5 GHz upper frequency covers Wi-Fi (2.4 / 5 / 6 GHz), Bluetooth, Zigbee, Sub-GHz IoT, cellular and common RF/microwave design frequencies.

PicoVNA 108 USB vector network analyzer: what it measures
The PicoVNA 108 captures both magnitude and phase across all four S-parameters in a single sweep, so one instrument handles return loss (S11, S22), insertion loss and gain (S21, S12), VSWR, impedance, group delay and time-domain reflectometry. Because measurement is vector, not scalar, it corrects systematic errors after calibration, recovering the 47 dB typical / 40 dB minimum corrected directivity and 46 dB typical / 40 dB minimum corrected source and load match that let it measure well-matched and high-rejection devices accurately. This is the core reason RF engineers choose a USB VNA over a simpler scalar analyzer.
PicoVNA 108 specifications
| Feature | Specification |
|---|---|
| Frequency range | 300 kHz to 8.5 GHz |
| Ports | 2 (full S-parameter: S11, S21, S12, S22) |
| Dynamic range | Up to 124 dB at 10 Hz |
| Trace noise | 0.006 dB rms (140 kHz bandwidth) |
| Directivity (corrected) | 47 dB typical / 40 dB minimum |
| Source match (corrected) | 46 dB typical / 40 dB minimum |
| Load match (corrected) | 46 dB typical / 40 dB minimum |
| Speed | >5000 dual-port S-parameters per second |
| IF bandwidth | 140 kHz to 10 Hz (selectable) |
| Frequency points | Up to 10,001 (VNA mode) |
| Test port connectors | Type N (female), 50 Ω |
| Output power | Up to +10 dBm (≤6 GHz), +6 dBm (>6 GHz), down to −20 dBm |
| Display formats | Log/linear amplitude, phase, group delay, VSWR, real, imaginary, Smith chart, polar, time domain |
| Calibration | 8- and 12-term: SOLT, unknown-thru, TRL, TRM; automated E-Cal available |
| Power | USB 2.0 controlled; external 12–15 V DC, 25 W (adaptor included) |
| Software | PicoVNA 5 (Windows 8/10/11 64-bit; macOS, Linux, Raspberry Pi) |
PicoVNA 108 vs PicoVNA 106: choosing the right USB VNA
Both PicoVNA instruments share the same compact design with external DC supply and USB 2.0 control, PicoVNA 5 software and calibration methods. They differ in upper frequency and dynamic range, so the choice comes down to how high in frequency you need to measure and how much rejection you need to resolve.
| Model | Frequency range | Dynamic range (at 10 Hz) |
|---|---|---|
| PicoVNA 108 | 300 kHz to 8.5 GHz | Up to 124 dB |
| PicoVNA 106 | 300 kHz to 6 GHz | Up to 118 dB |
The PicoVNA 108 is the right choice when your work extends above 6 GHz or when you need the extra dynamic range to characterize high-rejection filters and low-loss passives. GSAS helps you match the model to your highest measurement frequency with margin to spare.
Calibration and the automated E-Cal option
The PicoVNA 108 supports full 8- and 12-term error correction, SOLT, unknown-thru, TRL and TRM, using mechanical calibration standards, plus an automated E-Cal electronic calibration unit that is available for both the PicoVNA 106 and PicoVNA 108. E-Cal removes repeated manual standard swaps, shortens recalibration time and improves repeatability on the bench and on the line. GSAS advises on the calibration kit and connector standards that match your DUT.
Applications
The PicoVNA 108’s frequency reach, dynamic range and full vector correction make it well suited to:
- RF filter design and tuning: measuring passband insertion loss, stopband rejection and return loss with the dynamic range to see deep nulls.
- Antenna matching and characterization: return loss, VSWR and impedance for Wi-Fi, Bluetooth, IoT, cellular and sub-GHz antennas.
- Amplifier and attenuator test: gain (S21), input/output match and reverse isolation across the 8.5 GHz span.
- Cable, connector and PCB interconnect qualification: insertion loss, return loss and time-domain reflectometry to locate impedance discontinuities.
- Production and incoming inspection: fast, calibrated S-parameter screening with E-Cal for repeatable recalibration.
The PicoVNA 108 supports engineering and manufacturing teams in RF and wireless, aerospace and defence, IoT and connectivity, and electronics R&D and QA where accurate, traceable S-parameter measurement matters.
Watch the PicoVNA 108 introduction
VNA price in India: buy the PicoVNA 108 from GSAS Micro Systems
GSAS Micro Systems is an authorized Pico Technology engineering partner in India. We supply the PicoVNA 108 USB vector network analyzer with calibration-kit and E-Cal selection, precision RF accessories, hands-on demonstrations, GST-compliant quotations, calibration with maintained traceability and post-sales application support. RF, wireless and microwave engineering teams across Bengaluru, Hyderabad, Chennai, Pune, Mumbai and Delhi NCR rely on GSAS for fast access and local support. Contact GSAS or request a quote to get the PicoVNA 108 price in India and order with delivery nationwide.
Blog
Pico Technology Insights
How to Buy an Oscilloscope in India (2026): USB vs Benchtop, Bit Depth, Bandwidth: A Buying Guide
A vendor-neutral buying guide for Indian engineering teams choosing an oscilloscope in 2026. How to size bandwidth, sample rate, channels and bit depth; when a USB/PC oscilloscope beats a benchtop; and how oscilloscope price in India actually breaks down once you add GST, probes and warranty. Written for embedded, power and education teams in Bengaluru, Pune, Chennai, Hyderabad, Mumbai and Delhi NCR.
PicoScope 5000D vs 5000E: Native 16-Bit or FlexRes? A Buyer's Guide for Indian Labs
The established PicoScope 5000D FlexRes switches between 8-bit and 16-bit; the new PicoScope 5000E gives you native 16-bit always on. A verified, spec-by-spec buyer's guide for Indian labs choosing between them, with 5444D vs 5464E framing and India procurement notes from GSAS.
PicoScope 5000E in India: The World's First 4-Channel 16-Bit USB Oscilloscope, Explained
Pico Technology launched the PicoScope 5000E on 28 May 2026, the world's first 4-channel 16-bit USB oscilloscope. Verified specs, the 5462E/5463E/5464E and Plus models, and how to buy the 5000E in India through GSAS.
Related products
More from Pico Technology
All Pico Technology products →
PicoVNA 106
USB 2-port vector network analyzer covering 300 kHz to 6 GHz with up to 118 dB dynamic range for RF component and antenna characterization. Buy in India from GSAS with local support.
View PicoVNA 106 →
PicoSource AS108
USB-controlled agile synthesizer covering 300 kHz to 8 GHz with fast switching, IQ modulation, and programmable API. India pricing and local support from GSAS.
View PicoSource AS108 →
PicoSource PG900 Series
USB differential pulse generators with sub-60 ps rise time for TDR/TDT, jitter analysis, and high-speed interconnect testing. Local stock and support across India, GSAS.
View PicoSource PG900 Series →FAQ
Common questions about PicoVNA 108
How much does the picovna-108 cost in India?
What is the difference between the PicoVNA 108 and the PicoVNA 106?
Is the PicoVNA 108 a true vector network analyzer or a scalar analyzer?
What calibration does the PicoVNA 108 use, and is automated calibration available?
Which operating systems does the PicoVNA 5 software run on?
Is the PicoVNA 108 USB bus-powered?
How do I buy the PicoVNA 108 in India?
Interested in PicoVNA 108?
Get pricing, an evaluation unit, or a technical consultation from our application engineers.